Anesthesia Machine with Ventilator AM-AM32

Anesthesia Machine with Ventilator AM-AM32 offers multiple ventilation modes including IPPV, V-A/C, P-SIMV, etc for various patient needs. It provides a tidal volume (Vt) ranging from 0, 20 mL to 1500 mL. Coupled with a breathing circuit to deliver gases to the patient and remove exhaled gases smoothly. Equipped with a 10.4-inch TFT LCD screen for real-time monitoring of vital parameters and waveforms.

Catalog
Specifications
Ventilation Modes IPPV
V-A/C
P-A/C
V-SIMV
P-SIMV
PCV-PC
PCV-V
PSV
SIGH
MANUAL
Tidal Volume (Vt) 0, 20 mL to 1500 mL
Flowmeter O₂ (0.1 to 10 L/min)
N₂O (0.1 to 10 L/min)
AIR (0.1 to 10 L/min)
Pressure Limit 10 cmH₂O to 100 cmH₂O
Rapid Oxygen Supply 25 L/min to 75 L/min
Frequency (Freq) 1 /min to 100 /min
I: E 4:1 to 1:8
PEEP 0 cmH₂O to 30 cmH₂O
Pressure Triggering Sensitivity (PTr) 20 cmH₂O to 0 cmH₂O (Based on PEEP)
Flow Trigger Sensitivity (FTr) 0.5 L/min to 30 L/min
Pressure Control (PC) 5 cmH₂O to 70 cmH₂O
SIGH 0 (off) 1/100 to 5/100
Apnea Ventilation OFF, 5s to 60s
Oscillogram P-T (pressure-time)
F-T (flflow-time)
V-T (volume-time)
ETCO2-T (ETCO2-time)
P-V loop (pressure-volume loop)
P-F loop (pressure-flflow loop)
F-V loop (flflow-volume loop)
Low Battery Voltage Alarm < 10.2 ± 0.3 V
No Tidal Volume ≤ 5 mL within 6s
High Oxygen Concentration Alarm 19 % to 100 %
Low Oxygen Concentration Alarm 18 % to 99 %
High Airway Pressure Alarm 20 cmH₂O to 100 cmH₂O
Low Airway Pressure Alarm 0 cmH₂O to 20 cmH₂O
High-Minute Volume Alarm Adult (5 L/min to 20 L/min)
Low-Minute Volume Alarm Paed (1 L/min to 15 L/min, 0 to 10 L/min)
Continuous Pressure Alarm (PEEP +1.5 kPa) over 16 s
Suffocation Warning 5s to 60s no spontaneous ventilation
Maximum Limited Pressure ≤ 12.5 kPa
Gas Source 280 kPa to 600 kPa
Monitoring Frequency (Freq) 0 /min to 100 /min
Monitoring Tidal Volume (Vt) 0 mL to 2000 mL
MV 0 L/min to 100 L/min
AC Power Failure Alarm Power failure or no connection
Fan error Show on screen
Oxygen Deficit Show on screen
Oxygen Concentration 15 % to 100 %
Power input 80 VA
Power Consumption 100 to 240 V
Power Supply 50/60 Hz
Dimensions (L×W×H) 920×970×1380 mm
Net Weight 87.5 kg
Gross Weight 155 kg
Features
Optional Accessories

Vital sign monitor


AGSS
EtCO2 sensor

Applications

Anesthesia Machine with Ventilator offers advanced ventilation modes and ergonomic design for safe inhalation anesthesia and respiratory management in various medical settings, accommodating patients of all ages.

Anesthesia Machines

Anesthesia Machines are designed to ensure patient safety, efficiency, and reliability in every procedure. Engineered with precision, this cutting-edge device seamlessly integrates advanced technology for optimal anesthetic delivery. They offer Real-time analytics and alarms, minimizing risks and enhancing patient outcomes. Our compact device offers patient care and operational excellence with its smart performance, intuitive interface, comprehensive tracking, and durability.

Anesthesia Machine AM-AM31

Gas source O2: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min; 1.1 to 10L/min
Flow rate of rapid oxygen supply 25 to 75L/min

Anesthesia Machine with Ventilator AM-AM33

Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min N2O: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off
Anesthesia Machine with Ventilator AM-AM34
Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min N2O: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off
Anesthesia Machine with Ventilator AM-AM35
Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off