Anesthesia Machine AM-AM31

Anesthesia Machine AM-AM31 provided with latest refinements in ergonomics and systems integration with an advanced and easy to use anesthesia platform. Equipped with double-tube oxygen flow meter, round evaporator and standard loop- system. Main body is made up of high strength plastic and resistance towards chemical adhesives. High precision vaporizer is designed with a stable concentration unit.

Catalog
Specifications
Gas source O2: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min; 1.1 to 10L/min
Flow rate of rapid oxygen supply 25 to 75L/min
Vaporizer Among Halothane, En-flurane, Iso-flurane and Sevoflurane, one can be chosen for application as required by the customer
Respiratory circuit working mode: all-close, semi-close, semi-open releasing pressure: 1 to 4 Kpa
Respiratory bellows Bellows for adults, bellows for children Tidal volume range:0 to 1500 ml FM7150 Simple Anesthesia Machine Optional: N2O Gas
Features
Applications

It is widely used in hospitals, to deliver general anesthesia to patients as they undergo a medical procedure

Anesthesia Machines

Anesthesia Machines are designed to ensure patient safety, efficiency, and reliability in every procedure. Engineered with precision, this cutting-edge device seamlessly integrates advanced technology for optimal anesthetic delivery. They offer Real-time analytics and alarms, minimizing risks and enhancing patient outcomes. Our compact device offers patient care and operational excellence with its smart performance, intuitive interface, comprehensive tracking, and durability.

Anesthesia Machine with Ventilator AM-AM32

Ventilation Modes IPPV<br> V-A/C<br> P-A/C<br> V-SIMV<br> P-SIMV<br> PCV-PC<br> PCV-V<br> PSV<br> SIGH<br> MANUAL
Tidal Volume (Vt) 0, 20 mL to 1500 mL
Flowmeter O₂ (0.1 to 10 L/min)<br> N₂O (0.1 to 10 L/min)<br> AIR (0.1 to 10 L/min)

Anesthesia Machine with Ventilator AM-AM33

Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min N2O: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off
Anesthesia Machine with Ventilator AM-AM34
Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min N2O: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off
Anesthesia Machine with Ventilator AM-AM35
Gas source O2: 0.27 to 0.55MPa; N2O: 0.27 to 0.55MPa
Flow meter O2: 0.1 to 1.0L/min, 1.1 to 10L/min
O2, N2O linkage and N2O Stopper When using nitrous oxide, oxygen concentration>25%; When the oxygen pressure<0.2MPa, the flow of nitrous oxide would be cut off